BD LSRFortessa User Instructions
Hide thumbs Also See for LSRFortessa:

Advertisement

flowkon
Flow Cytometry User Instructions
at the University of Konstanz
FACSFortessa
The BD™ High Throughput Sampler (HTS) option for the BD LSRFortessa™ cell analyzer
provides rapid, fully automated sample acquisition from microtiter plates. In high-
throughput mode, the HTS can process a 96-well plate in fewer than 15 minutes.
Standard throughput mode can be selected for acquisition of larger sample volumes.

Preparing the HTS

Switch the acquisition control to plate mode.
This will switch off the DCM pump.
Unscrew the tube retainer (grey ring
DCM sleeve onto the SIT and carefully remove the sleeve.
Remove the DCM sleeve and replace it with the shorter SIT
protector.
Install the sample coupler on the SIT.
Check the level in the connected water bottle (> ~30 mL).
Switch on the HTS reader (on the backside).
Press the Run button (the HTS does not operate in STNDBY mode).

Initialisation and Priming the HTS

Note: Don't change PMTV of fluorescence channels after enabling compensation.
Operation of the loader is only possible via DIVA software – do not use the prime
button on the instrument.
When the BD DIVA software is already open, go HTS
Start BD DIVA software. The HTS will initialize automatically, otherwise go to HTS
Re-Initialize
Prime the HTS (Sheath tubing connected): HTS

Run your Experiment

See detailed instructions on HTS Programming.
– HTS User Instructions
TM
) that holds the
Re-Initialize.
Prime. 3x in total.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LSRFortessa and is the answer not in the manual?

Questions and answers

Summary of Contents for BD LSRFortessa

  • Page 1 FACSFortessa – HTS User Instructions The BD™ High Throughput Sampler (HTS) option for the BD LSRFortessa™ cell analyzer provides rapid, fully automated sample acquisition from microtiter plates. In high- throughput mode, the HTS can process a 96-well plate in fewer than 15 minutes.
  • Page 2 Cleaning Procedure after Measurement Include 3x4 wells for cleaning solutions on your plate (F9-F12, G9-G12, H9-H12)  OR use a separate plate for the cleaning procedure. Fill 250 µL FACSClean in 4 wells, F9-F12.  Fill 250 µL FACSRinse in 4 wells, G9-G12. ...

Table of Contents