Click the Next button and enter the high values as shown below.
Click Next the following window will be displayed showing the calibrated SYS value which is dependent on the
current input values.
The device is now calibrated. However you may find SYS has been 'clamped' if the resultant SYS is greater than
SMAX or less than SMIN. If this is the case then change these values to suitable limits. In this example we may set
SMIN to –0.5 (tonne) and SMAX to 12.0 (tonne). This would then provide clamping of SYS to these values and also a
flags being set in FLAG and STAT.

Sys Calibration, Auto Method

Assume we require to calibrate for Kg output and we have available a known accurate 10 Kg and 100 Kg test
weights.
19