Read/write Enumerated – These parameters can only be changed by
selecting the new value from a drop down list.
Clicking in the right hand column will display a down arrow button which
when clicked will display the parameter value options in a list.
Note that all enumerated data (apart from on/off) will be displayed with
a numeric value, hyphen then the description of the value.
The numeric value is the value of the parameter and the description is
just there to help.
Commands – These commands have 'Click to execute...' displayed in the
right hand column. Clicking here will display a
issue the command to the device.
As parameters are changed the communications traffic is displayed in the Traffic Pane.
If any errors occur these will be shown in red in the Error Pane. Once an error occurs it will need to be reset before
any more communications can take place. Reset errors by either right-clicking the Error Pane and selecting Reset
Errors from the pop-up menu or select the Communications menu and click the Reset Errors item.
To manually refresh the parameter list click the
menu.
Now you have successfully established communications with your evaluation device the next step is to perform a
simple calibration.

Connecting a Load Cell

You can now connect a strain gauge bridge, load cell or simulator to the DCell/DSC.
A suitable strain gauge should have an impedance of 350-5000ohms and (at least for now) a nominal output of
around 2.5mV/V.
15
Micron Meters
button. Click this to
button on the toolbar or select Sync Now from the Parameters
DCell & DSC User Manual