Download Print this page

Hach SC1000 Instruction Sheet page 2

Panel mount bracket
Hide thumbs Also See for SC1000:

Advertisement

sc1000 Panel Mount Bracket
Panel mount bracket installation
1
Bracket mounting holes (3x)
2
Pass-through hole, 26.746 mm (1.053 in.), for strain
relief (supplied)
2
1. Determine the location of the bracket on the panel. Allow enough room for the strain
relief pass-through or optional bulkhead connector pass-through to be located within a
228.6 mm (9 in.) radius of the bottom mounting screw
2. Install three customer-supplied # 10 pan head or equivalent low profile head screws
into the panel
(Figure
3. Hang the sc1000 panel mount bracket over the screws.
4. Hang the sc1000 display module on the panel mount bracket
5. Drill the pass-through hole for the strain relief fitting or the optional bulkhead
connector within a 9-inch radius from the bottom mounting screw location
Figure 1 Panel mount dimensions
1, item 1).
3
Maximum 9-inch radius for pass-through holes
4
Pass-through hole, 12.319 ± 0.381 mm
(0.485 ± 0.015 in.) for optional bulkhead connector
(Cat. No. 71211-00, not supplied)
(Figure
1).
(Figure
2, item 8).
(Figure
1).

Advertisement

loading