Servicing - KNF NF 10 DC series Operating Instructions Manual

Diaphragm liquid pump
Table of Contents

Advertisement

Diaphragm–Liquid pump NF 10
Information on procedure
Prior requirements
KNF Flodos AG BA_NF10_EN_05_069219
Operating and Installation Manual
8.

Servicing

8.1. Servicing schedule
Component
Pump
Pump head
Diaphragm and valves
Tab. 14
8.2. Cleaning
WARNUNG
8.2.1. Flushing the pump
 When pumping aggressive media, KNF recommends flushing
the pump with a neutral fluid under atmospheric conditions for
a few minutes before switching off, to extend the service life of
valves and diaphragm.
8.2.2. Cleaning the pump
 Wipe the outside of the pump with a soft dry cloth. Ensure that
any cleaning solvents used will not damage the materials of
the pump head and crankcase housing (check material
resistance properties).
 If there is compressed air available, blow off the separate
parts.
Pump must be switched off and mains plug removed from
the socket.
The pump must be free of any hazardous substances.
Hoses must be disconnected from the pump head.
It is recommended to replace the diaphragm when the
head parts are removed.
Tools
Qty.
Tool
1
Screwdriver TORX T6
Tab. 15
Servicing interval
-
Regular inspection for external
damage or leaks
-
Clean the pump internally: if the
flow rate changes, the pump does
not pump or no vacuum is created
(section 8.2)
-
At the latest change when the flow
drops off
Health hazard due to dangerous substances
in the pump!
Depending on the substance transferred, caustic
burns or poisoning are possible
 Wear protective clothing if necessary, e.g. pro-
tective gloves.
 Rinse the pump with a neutral liquid and pump
empty.
Servicing
18