during calibration, the
added
weights
ERR2
enough
during calibration, input
single is negative.
ERR3
During
calibration,
ERR4
single is unstable
ERR5
EEPROM check error
6.2
Daily maintenance
1. In order to ensure indicator display clearly and prolong use life, the
indicator should not be placed directly on sunlight.
2. Load cell and indicator should be well connected , the system should
have a good ground, away from strong electric field, magnetic field.
3. Do not use indicator outside in rainy, better keep it power off.
4. Power off firstly while plug and unplug
Page 25 of 27
User Manual
Added
weight
at
least
not
10%of
Max.
capacity,
Recommend the weights is
60-80% the Max. capacity
1..Check
connection
is
correct or not.
Check load cell is damaged
or not.
3. renew calibration, if still
wrong. pls replace the PCB
Ensure added weight and
scale
is
stable,
start
calibration
change PCB.