C04 Maximum capacity
For example:max weighing 100kg:
set [0100.00]
Press
, go to next step
C05 Zero calibration
option:0=non-calibration zero
calibration zero please choose 1 and ensure scale is empty and "stable" light
is on. countdown[CAL 10]~[CAL
show[0.00](example for two decimal point).
C06 Loading calibration
[C06], Press
, show [C06
Press
, show [SPAn ],
Basic on max capacity setting, add suitable weight on scale. close to the max
capacity, heavier than 10% max at least.
For example: the weight is 80kg
Page 15 of 27
User Manual
1=need calibration zero
0], then the indicator will
0], press
, change to [C06