Power Supply - GE 1926A-1AN Technical Manual

19 inch color lcd, resistive-touch
Table of Contents

Advertisement

● POWER SUPPLY. The device must be connected to a properly installed power outlet
with protective earth contacts only. If the integrity of the protective earth conductors is in
doubt, disconnect the LCD Display from the power line and use it with the battery option if
available). If the installation does not provide for a protective earth conductor, disconnect
the LCD Display from the power line. All devices in a system must be connected to the
same power supply circuit. Devices which are not connected to the same circuit must be
electrically isolated when operated.
● GE is responsible for the effects on safety, readability, and performance of the equipment
only if:
- Assembly operations, extensions, readjustments, modifications, servicing, or repairs are
carried out by authorized service personnel.
- The electrical installation of the relevant room complies with the requirements of the
appropriate regulations.
- The equipment is used in accordance with the instructions for use of the patient display
and this Technical manual.
● Grounding reliability can only be achieved when the equipment is connected to an
equivalent receptacle marked "Hospital Only" or "Hospital Grade".
● DO NOT position the LCD Display so that it is difficult to operate or to connect and
disconnect the AC power cord.
●DO NOT touch the patient simultaneously with connection or disconnection of cables.
●Installation and OSD adjusting should only be carried by manufacturer trained and
authorized personnel.