Recommended Separation Distances - GE 1926A-1AN Technical Manual

19 inch color lcd, resistive-touch
Table of Contents

Advertisement

Recommended separation distances between portable and mobile RF communications
equipment and the LCD display for Healthcare Applications for all EQUIPTMENT AND
The LCD display for Healthcare Applications is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the user of the
LCD display for Healthcare Applications can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications
(equipment) and the LCD display for Healthcare Applications as recommended below
according to the maximum output power of the communications equipment.
Rated Maximum
Output Power of
Transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distanced in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
14
SYSTEMS that are not LIFE-SUPPORTING
Separation Distance According to Frequency of Transmitter
150 kHz to 80 MHz
0.12
0.37
1.2
3.7
12
80MHzto800MHz
0.12
0.37
1.2
3.7
12
800MHz to 2.5 GHz
0.23
0.74
2.3
7.4
23