Loss/Length Testing in Smart Remote Mode (cont.)
6
Clean the tester's and source's OUTPUT connectors and
the reference test cord connectors.
7
On the main tester: press
Length Reference.
8
Connect the reference test cords as shown on the screen
and in Figure 28; then press
68
F
; then select Set Loss/
t
.
9
Optional: From the TEST SETUP screen, you can enter
reference test cord lengths to meet TSB-140 reporting
requirements:
L
Use
to highlight a reference test cord number, press
t
; then enter a length. Press
done.
C
10 Press
OK to leave the TEST SETUP screen.
W
If you disconnect the reference test cords from
the tester's or remote's output, you must set the
reference again to ensure valid measurements.