7
Transmitter keys, operation modes and operator menus
Front panel keys
The transmitter is operated using the keys on the front panel. Prompts associated with active keys are displayed on each screen.
Diagnostic messages are detailed on page 15, display icon descriptions are detailed on page 17.
Fig. 6 Front panel keys
Key
Function
A
Navigation key – left
and
Operator menu access key
B
View key
C
Up key
D
Down key
E
Group key
F
Navigation key – right
and
Cal shortcut key
Table 1 Key functions
Operator
Page 1
Diagnostics
View
Calibration
Log
Fig. 7 Menu navigation overview
CI/AWT440–EN Rev. A
Description
When any Operating, View or Log page is displayed, opens or closes the Operator menu
and returns to the previous menu level.
Toggles the view between Operator pages, Diagnostic View and Calibration Log screens
– see Fig. 7.
Note. Disabled in Configuration mode.
Used to navigate up menu lists, highlight menu items and increase displayed values.
Used to navigate down menu lists, highlight menu items and decrease displayed values.
Toggles between:
— Operator pages (1 to 5) when an Operator page is selected with the View key.
— View screens (Diagnostics, Signals, Alarms, and Outputs) when the Diagnostic
View screen is selected with the View key.
— Log screens (Calibration, Alarm, Audit and Diagnostic) when the Calibration Log
screen is selected with the View key.
See Fig. 7.
Note. Disabled in Configuration mode.
At menu level, selects the highlighted menu item, operation button or edits a selection.
When any Operating, View or Log page is displayed, used as a shortcut key to access
the Calibrate level.
Operator
Operator
Page 2
Page 3
Signals
Chart
View
View
Alarm
Audit
Log
Log
Operator
Operator
Page 4
Page 5
Alarms
Outputs
View
View
Diagnostic
Log
11