Technical Data Sheet - Miele IR6002 Installations Plan

Laboratory programmable washer
Hide thumbs Also See for IR6002:

Advertisement

Available languages

Available languages

Technical Datasheet
Industrial cleaner
Legend:
The unit should only be transported by a palletiser or fork lift on Miele pallet!
Electrical
connection
Country variations:
Plug and socket connection recommended to facilitate
electrical safety tests. Socket must be accessible after
installation.
Connection to electrical supply should be in accordance with
valid regulations and safety standards. Observe correct
phase sequence (clockwise). Protect supply lead against
thermal exposure.
The use of a current-operated ELCB is recommended!
Cold water
Max. temperature
Max. water hardness
Min. flow pressure
Min. flow pressure with extended intake time
Max. pressure
Throughput
On-site connection thread according to DIN 44 991
Length of connection hose (supplied)
DVGW approved
Direct connection without non-return valve permissible.
Installationsplan IR6002
Stand: 01.12.2007
Circled, bold-type abbreviations:
Connection required
Abbreviations surrounded by broken circle:
Connection optional or required, depending on model
1. Voltage
Rated load
Fuse rating
Connection cable, min. cross-section
Length of supply lead (H05(07)RN-F) without plug
2. Voltage
Rated load
Fuse rating
Connection cable, min. cross-section
Length of supply lead with plug
3. Voltage
Rated load
Fuse rating
Connection cable, min. cross-section
Length of supply lead with plug
3. Voltage
Rated load
Fuse rating
Connection cable, min. cross-section
Length of supply lead with plug
IR 6002
V/Hz
3N AC 400/50
kW
10.2
A
3 × 16
mm²
5 × 2.5
m
1.5
V/Hz
3N AC 400/50
kW
10.2
A
3 × 15/16
mm²
5 × 2.5
m
1.5
V/Hz
3 AC 208/60
kW
6.6
A
3 × 20
AWG
4 × 12
m
1.9
V/Hz
3N AC 380/60
kW
9.3
A
3 × 16
AWG
5 × 2.5
m
1.5
° C
20
° dH
60
kPa
200
kPa
30
kPa
1000
l/min
10.5
Inch
3/4" external thread
(USA: 11.5 NH)
mm
1500
Seite 5