Miele IR6002 Installations Plan page 9

Laboratory programmable washer
Hide thumbs Also See for IR6002:

Advertisement

Available languages

Available languages

Hot water
Max. temperature
Max. water hardness
Min. flow pressure
Min. flow pressure with extended intake time
Max. pressure
Throughput
On-site connection thread according to DIN 44 991
Length of connection hose (supplied)
DVGW approved
Direct connection without non-return valve permissible.
De-mineralised
The quality (conductivity) of the water used is defined by rinse
water
performance requirements.
Min. flow pressure
Min. flow pressure with extended intake time
Max. pressure
Throughput
Connection thread (on site)
Length of connection hose (supplied)
Waste water
Drain hose (Int. dia. × wall thickness × l)
Max. head height (measured from floor)
Max. transient throughput
Hose sleeve for drain hose to be provided on site
(ext. dia. x l)
Recycle drainage
Connection only required if process water is to be recycled
(water discharged from various programmes can be collected
separately in on-site tank).
If the IR 6002 is not connected to an on-site tank and the
DUO drainage function is selected, both drain hoses must be
vented.
In this case, remove the blind stopper from the recycling
drainage connection and attach the vent hoses supplied
using the enclosed Y-piece to the standard hose vent.
Max. temperature
Max. peak delivery
Max. head height (measured from floor)
Drain hose (Int. dia. × wall thickness × l)
Length of connection hose (supplied)
Hose sleeve for drain hose to be provided on site
(ext. dia. x l)
Recycle Intake
Connection of recycling water intake (water from on-site tank)
Machine-side connection (ext. dia. × length)
(Connection hose to be provided on site)
The tank is installed next to the industrial cleaner at a height
of at least 450 mm above finished floor level. The tank must
be vented and have an overflow pipe at a max. height of 950
mm. The tank must have a capacity of 20 litres at the
overflow level. The inlet hose should be installed at a steady
incline from the tank to the machine connection point.
Precautions must be taken to prevent foreign matter from
entering the machine via the tank. If required, cover the tank
with a lid and fit a strainer.
Installationsplan IR6002
Stand: 01.12.2007
° C
70
° dH
60
kPa
200
kPa
30
kPa
1000
l/min
10.5
Inch
3/4" external thread
(USA: 11.5 NH)
mm
1500
kPa
150
kPa
30
kPa
1000
l/min
10
Inch
¾ external
(USA: 11.5 NH)
mm
1500
mm
22 × 6 × 1500
m
1.0
l/min
50
mm
22 × 30
° C
93
l/min
50
m
1.0
mm
22 × 6 × 1500
mm
1500
mm
22 × 30
mm
25 × 30
Seite 6