Vent connection
Vented to atmosphere without fan:
Venting is provided by a vent hose supplied with the machine
Connection of vent hose to vent ducting via supplied hose
connector (outer diamter x length)
Max. permissible length of vent duct using HT pipe DN 70
Reduction in max. permissible vent duct length for each 90°
bend.
Temperature - Mean value / peak max.
Rel. humidity - Mean value / peak max.
Vent air throughput
Sufficient air intake should be available to replace displaced
volume of air. Do not merge vent ducting from multiple
machine!
Prevent condensate backflow into washer-disinfector. Pitch
vent ducting and discharge condensate at lowest point.
Dispensing
Direct connection of two (four optional) siphons to supply
canisters is possible. Canister (to be provided on site) can be
installed to the left or right of the machine.
Machine data
Height incl. lid
Width
Depth
Net weight
Dynamic floor load
Knocked down width incl. pallet
Knocked down height incl. pallet
Installation should only be carried out by authorised fitters in accordance with valid regulations!
Observe installation instructions when installing machine! All rights reserved! Measurements in mm
Installationsplan IR6002
Stand: 01.12.2007
mm
36 × 30
m
5.0
m
0.5
° C
70/95
%
80/100
m³/h
100
mm
1175
mm
900
mm
700
kg
200
N
3000
mm
800
mm
1400