Kusam-meco KM-CAL-804 Operation Manual

Kusam-meco KM-CAL-804 Operation Manual

Voltage / ma calibrator

Advertisement

Quick Links

An ISO 9001:2008 Company
The new KUSAM-MECO Volt & milliampere Calibrator Model KM-CAL-804 is a high
accuracy (0.05%) Calibrator. It has 5 digit LCD display with high stability. It can be used
for calibrating Transmitters, Thermocouples, SCADA systems, Transducers.
provides loop circuit power 24V.
The max. load is 1kW at 20mA range when the power is higher than 6.8V; The max. load
is 700W at 20mA range when the power is within 5.8V - 6.8V. Temperature Coefficient
± 0.005%, range / °C (5°C -18°C, 28°C - 40°C).
battery life is approx 20 hours under the condition of 10mA.
GENERAL SPECIFICATIONS
í Basic Accuracy : ± 0.05%
í Display : 5 Digit LCD display.
í Max. Allowable Voltage : 30V..
í Operating Temperature Range : 0 ~ 50°C.
Humidity range : £ 80% RH
í Storage Temperature Range : £ -10°C ~ 55°C
Humidity range : £ 90% RH
SAFETY :
Complies with IEC1010 (safety standard issued
by International Electrician Committee)
OUTPUT FUNCTION
Output
Range
DC Voltage
10V
DC Current
20 mA
Simulate
transmitter
-20 mA
(sinking Current)
Loop Power
24 V
®
An ISO 9001:2008 Company
Chhaya com/D/chhaya/my documents/chhaya/backup/catlog/New catlog/2015 new arrival catalogs/KM-CAL-804.cdr
VOLTAGE / mA CALIBRATOR
It works on 9Volt alkaline battery and
í Temperature Coefficient : 0.1 x (dedicated Accuracy) % /
°C (5°C ~ 18°C, 28°C ~ 40°C)
í Power : 1.5V x 2 alkaline batteries.
í Power Consumption : about 400mA / 3V in condition of
1kW load & 20mA output.
í Dimension : 180 (L) x 90 (W) x 47 (D) mm (with protector)
í Weight : About 500g
ACCESSORIES :
User Manual, Test lead CF-36, (Clips for probe),
Holster & Carrying case.
ELECTRICAL SPECIFICATIONS - KM-CAL-804
Output Range
Resolution
0.000 ~ 11.000V
0.001V
0.000 ~ 22.000mA
0.001mA
0.000 ~ -22.000mA
0.001mA
G-17, Bharat Industrial Estate, T. J. Road, Sewree (W), Mumbai - 400 015. INDIA.
Sales Direct.: 022-24156638,
Email : kusam_meco@vsnl.net,
Model KM-CAL-804
It also
Accuracy
± 0.05% + 2
± 0.05% + 4mA
± 0.05% + 4mA
± 10%
All Specifications are subject to change without prior notice.
Tel. : 022-24124540, 24181649,
Website : www.kusamelectrical.com
Preliminary Data
Remarks
Max. Output
current 5mA
20mA Max. load 1KW
20mA Max. load 1KW
Note:
power supply range :
5 ~ 25VDC
Max. Output
current 25mA
Fax : 022-24149659

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the KM-CAL-804 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Kusam-meco KM-CAL-804

  • Page 1 Model KM-CAL-804 An ISO 9001:2008 Company The new KUSAM-MECO Volt & milliampere Calibrator Model KM-CAL-804 is a high accuracy (0.05%) Calibrator. It has 5 digit LCD display with high stability. It can be used for calibrating Transmitters, Thermocouples, SCADA systems, Transducers.
  • Page 2 Digital Hand Held Temperature Indicators ® G 17, Bharat Industrial Estate, T. J. Road, Sewree (W), Mumbai - 400 015. INDIA. Sales Direct : (022) 24156638 OPERATION MANUAL Tel. : (022) 24124540, 24181649. Fax : (022) 24149659 Email : kusam_meco@vsnl.net Website : www.kusamelectrical.com www.kusam-meco.co.in...
  • Page 3 ® ® TABLE OF CONTENTS Section One Safety Information To ensure safe use ,the meter and manual employ the following symbols: TITLE PAGE NO. Warning identifies conditions and actions that may pose Safaety Information ........hazard (s) to the user and avoid methods. Instrument Panel Layout and Functions..
  • Page 4: Section Three Maintenance

    ® ® Note OUTPUT : indicates the meter is in output state. • To ensure accuracy, preheat for 5 minutes after power on, CAL : indicates the meter is in calibration state. • Please contact the manufacture or dealers if the Users have higher accuracy requirement.
  • Page 5 ® ® 1 ) Turn the meter off and remove all test leads. 2 ) Shake out any dirt that may be in terminals. 3 ) Soak a new swab with alcohol. Clean each terminal with the swab. (2) Replacing the batteries The meter is powered by two LR6 alkaline batteries (AA).
  • Page 6 ® ® Section Five Output of the Meter The meter generates DC current or simulate resistance set by the Users from the corresponding output terminals ( OUTPUT) . Caution Do not apply any voltage to output terminal; otherwise damage to interior circuit may occur if the voltage is not proper.
  • Page 7 ® ® 4. Press( ) / ( ) key, select output set bit; 5. Press ( A ) / (V) key , change the value of set bit, and the value can carry or abdicate automatically, and hold the key, the value will alter constantly after one second.
  • Page 8: Section Six Function Setting

    ® ® Section Six Function Setting 4. Press ( ) key to set selection as 100% , and the current outputs ' 20mA . Press ( ) key to set selection as 0% , and the current outputs The following operation can change the automatic power off 4mA) 5.
  • Page 9 ® ® General Feature Section Eight Note for the Manual • • power : two 1.5V alkaline batteries(LR6) The present operation instruction is subject to change without notice; • power consumption : about 400mA /3V, in condition of 1k load and •...
  • Page 10: Test Certificate

    ® ® WARRANTY MUMBAI Each “KUSAM-MECO” product is warranted to be free from defects in material and workmanship under normal use & service. The TEST CERTIFICATE warranty period is one year (12 months) and begins from the date of despatch of goods. In case any defect occurs in functioning of the...