Technical Information
Inlet screen installation
prevent the blades of the pump from being damaged by debris. The
inlet screen, however, does reduce the pumping speed as followed:
X3500-68000
X3500-68001
The inlet screen is fitted in the upper part of the pump, as shown in
the figure.
Figure 36
The screen can be mounted on each pump.
The screen can be removed as shown in the following figure.
274/304
969-9304 (CFF 8'')
Reduction in pumping speed (%)
Ar
20%
N
2
18%
TwisTorr 304 FS User Manual / 87-901-027-01
He
H
2
7%
5%