Sense2 Subsystem (Option Process Interface) - Fluke Norma 4000 User Manual

Guide for remote control for power analyzer
Hide thumbs Also See for Norma 4000:
Table of Contents

Advertisement

Invalidated by
-
[SENSe:]SWEep1|2:FUNCtion <function>{,<function>}
Description
Specifies function list for memory recording. The maximum number of functions is 20. If
the list exceeds device capability, the sample factor is automatically increased.
For SWEep1 (REALtime) function list only sampled values can be specified.
For SWEep2 (AVERage) function list any functions from the standard instrument
function list can be specified.
The function list is stored in a saved configuration and is reloaded at PowerOn or with
*RCL.
Parameter(s)
<function>
SWEep1 (REALtime) valid functions:
VOLTage1..6[:DC]
CURRent1..6[:DC]
POWer1..6[:ACTive]
TORQue[1..4]
SPEed[1..4]
POWer[1..4]:MECHanical
SWEep2 (AVERage) valid functions:
Any functions from the standard instrument function list can be specified.
*RST state
"VOLTage1"
Example(s)
SWE1:FUNC "VOLT1", VOLT2","VOLT3"
Invalidates
-
Invalidated by
-

SENSe2 Subsystem (Option Process Interface)

The SENSe2 subsystem controls the settings of the optional Process Interface inputs.
Numeric suffixes of the TORQue|SPEed|POLepairs|TYPe|REFerence nodes correspond
to the index of the 4 motors/generators supported. Channels of the INPut subsystem are
combined into drive index of the SENSe2:xxx subsystem as shown in the table below.
Remote Control - Description of Commands
Commands and Queries
4
4-51