Remote Control
Users Guide
*RST state
0.3 s
Invalidates
SYNC:TIMeout
Invalidated by
-
Sampling Frequency
[SENSe:]SWEep:FREQuency?
Description
Queries the sample rate of the instrument's ADCs. The sample rate is fixed and cannot be
changed.
Response
<sample_rate>
The actual sample rate the instrument is using to acquire the data. The sample rate is
common to all channels. The unit is Hz.
*RST state
Device dependent:
Norma3000: 102.4 kHz
Norma4000: 341.33 kHz or 1.024 MHz
Norma5000: 341.33 kHz or 1.024 MHz
Example(s)
SWEep:FREQ?
Invalidates
-
Invalidated by
-
Measurement Functions
[SENSe:]FUNCtion[:ON] <function>{,<function>}
Description
The FUNCtion[:ON] command selects the <function>(s) to be SENSed by the instrument.
The <function> is specified as a quoted string. For example: FUNCtion "VOLTage:AC".
If CONCurrent is OFF, a single <function> is sent as the parameter. This selects this
function to be sensed. If more than one function is sent an error -108 (Parameter not
allowed) is generated. If CONCurrent is ON, a comma separated list of
<sensor_function> may be sent as parameters. These functions are turned on, while the
state of other functions is changed to off. The query response of FUNCtion[:ON]? returns
a comma separated list of functions that are on, each of which are <STRING RESPONSE
DATA>. If no functions are ON, a null string is returned. The query returns the short
form mnemonics and omits any default nodes in the <function>.
This function list does not get saved with *SAV and is cleared with *RST.
4-42
Response: 3.4133E+05