Not enough turns
Target not locked
Sensor specific messages Subsystem: FI_Darwin_Drives
Value
Value description
Position HV Response
A
Position HV event
B
Target
C
Additional results
perfoming a
Diagnostics:
Value
Blocked
Notifications
0
Performance
Item
System Monitor messages
and message codes stored in
*.fsmx file
1
Subsystem error drives
Drive system not initialied 2
3 6 B D D1
Focus 30/35 Service Manual
Status
IM presets H/V
Ok
position, where to drive.
H/V position reached
Ok
and positive event to IM.
Tracker has locked to
Ok
target.
D1
C1
Unblocked
0
Error Log stored in
*.irep file
FI: 10 Code: 0xf005
Message:
3|0|99|0|1104500|PositionHV
faile
FI: 10 Code: 0x9c52
Message: Req x6e got res 9
FI: 1 Code: 0x9c4e Message:
Drive system: Not initialized
( 2 3 6 B D
4 - 32
Troubleshooting
Ok Counter
1
1
1
Pos-
Stopped
Reached
0
1
Recommendation
Check Fine drive vertical
Exchange the Fine drive
vertical
P/N 79000035_SM, Version 1.5
Chapter 4
Error Counter
- Ok status
0
- Error message
1
- Ok status
0
- Error message
1
- Ok status
0
- Error message
1
Could NOT turn 4
times complete to
position.
Could NOT lock with
Tracker to prism
target.
Failure
0
Need help?