Changing Units; Calibrating The Gauge; Taking Measurements; Changing The Parameter Settings - Extech Instruments TKG100 Quick Start Manual

Ultrasonic thickness gauges
Hide thumbs Also See for TKG100:
Table of Contents

Advertisement

Extech Instruments 
 
4. Changing Units 
To change units from IN (inches) to MM (millimeters) or 
the probe is selected, press F1 Menu, scroll to initial settings, press Menu/OK, scroll to units, press 
Menu/OK, choose the desired unit, and then press Menu/OK and then F1 (Back). 
5. Calibrating the Gauge 
Calibrating is the process of adjusting the Gauge for a specific material and transducer before testing the 
material to ensure that all measurements are accurate. Always perform a calibration before measuring 
material to obtain standard accuracy. The following steps describe how to perform a Velocity and Zero 
Calibration. A test step block of known thickness is required to perform an accurate calibration.  
Velocity and Zero Calibration 
Note: it is highly recommended that a test block with known thicknesses and multiple steps of the 
same material with flat and parallel surfaces be used for calibration. 
To calibrate both Zero and Velocity at the same time (strongly recommended) first go to the 
Calibration mode by pressing menu, calibration, and OK. Messages will appear on the bottom of the 
display prompting the user through the process "couple thin, press zero..." While measuring the 
thinnest step, select Zero by pressing F2. After selecting Zero the transducer can be taken off of the 
test block. If the displayed value is different than the known value of the step, adjust the value using 
the up and down arrow keys and then press F1 (CAL). Then, while measuring the thickest step, select 
'VEl' by pressing F3. After selecting ZERO the transducer can be taken off of the test block. If the 
measured value is different than the known value of the step, adjust the measured value using the up 
or down arrow key and then press OK. The unit will briefly display the acoustic sound velocity value 
and then return to Measure mode. Note that the order of Zero and Velocity calibration could be 
reversed as long as Zero is used on the thin section and Velocity on the thick section.  
6. Taking Measurements 
Once an Auto‐zero is performed on the selected transducer, the unit automatically goes to the measure 
mode. To take thickness readings, the user can simply apply the (fluid) couplant to the surface and put 
the transducer on the test piece, applying some pressure, to measure thickness. Depending on the 
parameter settings, the display may show other parameters. For more detailed information on Measure 
mode, please refer to the User's Guide located on the supplied CD‐ROM.  
7. Changing the Parameter settings 
To change any parameter setting, press Menu/OK and then go to Measurement by pressing the up or 
down arrow key and then Menu/OK. The Measurement screen lists the parameters in alphabetical order 
such as alarm, diff, E to E, fast...these vary based on which Gauge is purchased. The procedure to change 
any parameter is the same...scroll to the parameter, press Menu/OK, scroll to select, and then press 
Menu/OK. Use the F1 key as the 'back' key to return to previous screens and main measurement modes.  
Copyright © 2013 FLIR Systems, Inc. 
All rights reserved including the right of reproduction in whole or in part in any form 
www.extech.com 
sec (microseconds): At the first screen, before 
TKG Series Quick Start Guide ‐ EN v3.4   11/13 