Download Print this page
SciCan Hydrim C51w Quick Reference Manual
Hide thumbs Also See for Hydrim C51w:

Advertisement

Hydrim
For best results:
Always use the heavy duty cycle for cassettes, hinged instruments,
heavily soiled instruments and the cannula post adapter.
Do not allow instruments to dry for more than one hour prior to
processing in the Hydrim.
Remove all cement, composite material and amalgam at chairside prior
to processing in the Hydrim.
Do not overload cassettes or baskets with instruments. (Follow the
manufacturer's guidelines.)
Place hinged instruments in baskets, not cassettes.
Do not use any detergent or cleaning agent other than HIP
Hydrim Cleaning Solution. It is not necessary to soak instruments
or cassettes in other detergents prior to placing in the Hydrim.
Place cassettes as indicated, to ensure that fluid fully contacts
instruments and cassettes.
Caution:
1.
HIP Cleaning Solution (Hydrim Cleaning Solution with Instrument
Protection) has been carefully formulated to provide
outstanding material compatibility with a wide range of metals
and coatings. (Some manufacturers' instruments may not be compatible.)
2.
Plastic objects that are not resistant to hot water (65˚C / 149˚F) may not be
suitable.
3.
When processing small objects, use the hinged basket (01-108294).
Failure to do so will result in small objects being thrown out of the baskets,
which may cause damage to the machine.
Description
Temperature
(wash)
Wash / Rinse
Time (min.)
Water
Consumption
Art w ork Num ber : 0 9-1090 81 R e v. 1.0
C51w
Quick Reference Guide
®
Art w ork Name: Ar t work, fa scia , H ydr im Dry
Art w ork cre ate d: Ju n e 29, 20
Par t N u mb er 92-109 082 R ev 1.0
Par t N a me: Lab el, fa scia, Hy dri m Dr y
PANTONE 624 C
PANTONE
Cool Grey 5 C
Black
Rinse
Cycle
and Hold
Use to prevent soil
from drying on
instruments and if
wash cycle is not
immediately desired.
35˚C / 95˚F
8 min.
6 litres /
1.59 gallons
05
Par t cr
eat ed b y G. G.I
Rel e as ed b y EC O 0 5-23 6
August 200 5
Cycle Complete
Regular Wash &
Dry
(approx.19min)
Heavy Duty
Wash & Dry
(approx. 22 min)
Rinse & Hold
(approx. 8 min)
(Cutouts)
(Windows)
Solution Low
Regular
Heavy Duty
Wash
Wash
Use for
Use for heavily
moderately
soiled instruments
soiled loose
or cassettes or
instruments.
the cannula post.
50˚C / 122˚F
50˚C / 122˚F
19 min.
22 min.
17 litres /
17 litres /
4.5 gallons
4.5 gallons
Indicator
Cycle Complete
Regular
Wash & Dry
Heavy Duty
Wash & Dry
Rinse &
Hold
Cleaning
Fault / Error
Salt
Salt Level
Window
Chem. Error

Advertisement

loading

Summary of Contents for SciCan Hydrim C51w

  • Page 1 Art w ork Num ber : 0 9-1090 81 R e v. 1.0 Hydrim C51w Quick Reference Guide ® Art w ork Name: Ar t work, fa scia , H ydr im Dry Art w ork cre ate d: Ju n e 29, 20 For best results: Par t N u mb er 92-109 082 R ev 1.0 Always use the heavy duty cycle for cassettes, hinged instruments,...
  • Page 2 Fill the water softening salt container with water softening salt. Screw the container back into position. For all service and repair inquiries: United States: 1-800-572-1211, International:49-(0) 821 - 278 93400, Canada: 1-800-870-7777 Email: techservice.ca@scican.com Hydrim is a registered trademark of SciCan, Division of Lux & Zwingenberger Ltd.