Nutricia Flocare 800 Short Manual

Enteral feeding pump

Advertisement

Quick Links

Flocare 800
Short Manual

Preparation and start of the pump

800
ENTERAL FEEDING PUMP
RATE ML/HR
RUN
HOLD
DOSE COMPLETE
VOL
OCCLUSION
DOSE
FED
NO FLOW
LOW BATTERY
DEC
INC
MALFUNCTION
AC POWER
> Insert the drip chamber of the feeding set into the drip detector slot of the pump.
1
> Slowly open the roller clamp on the feeding set and allow the total set to fill with feed. Make sure the drip chamber
2
is filled up about one-third.
3
> Stretch the tubing around the pump wheel and insert the plastic retainer into the bracket on the pump.
> Pull the tubing through the tubing guide.
4
ENTERAL FEEDING PUMP
RATE ML/HR
RUN
Connect the feeding set to your tube and continue to operate the pump.
DOSE COMPLETE
VOL
OCCLUSION
FED
NO FLOW

Operating instructions

LOW BATTERY
ENTERAL FEEDING PUMP
ENTERAL FEEDING PUMP
INC
MALFUNCTION
RATE ML/HR
RATE ML/HR
RUN
Turning on
AC POWER
DOSE COMPLETE
DOSE COMPLETE
the pump
VOL
OCCLUSION
OCCLUSION
FED
NO FLOW
NO FLOW
LOW BATTERY
LOW BATTERY

Programming

INC
ENTERAL FEEDING PUMP
MALFUNCTION
MALFUNCTION

Continuous Feeding

ENTERAL FEEDING PUMP
ENTERAL FEEDING PUMP
AC POWER
AC POWER
ENTERAL FEEDING PUMP
RATE ML/HR
RATE ML/HR
RATE ML/HR
RUN
DOSE COMPLETE
RATE ML/HR
DOSE COMPLETE
DOSE COMPLETE
Programming
OCCLUSION
VOL
DOSE COMPLETE
OCCLUSION
OCCLUSION
NO FLOW
FED

Dose Feeding

NO FLOW
NO FLOW
OCCLUSION
LOW BATTERY
NO FLOW
LOW BATTERY
LOW BATTERY
ENTERAL FEEDING PUMP
MALFUNCTION
INC
LOW BATTERY
MALFUNCTION
MALFUNCTION
RATE ML/HR
AC POWER
MALFUNCTION
ENTERAL FEEDING PUMP
AC POWER
AC POWER
DOSE COMPLETE
AC POWER
Changing
RATE ML/HR
OCCLUSION
ENTERAL FEEDING PUMP
NO FLOW
Programming
ENTERAL FEEDING PUMP
ENTERAL FEEDING PUMP
DOSE COMPLETE
LOW BATTERY
RATE ML/HR
OCCLUSION
RATE ML/HR
RATE ML/HR
NO FLOW
MALFUNCTION
DOSE COMPLETE
LOW BATTERY
DOSE COMPLETE
OCCLUSION
DOSE COMPLETE
AC POWER
Viewing total volume
NO FLOW
OCCLUSION
OCCLUSION
MALFUNCTION
NO FLOW
NO FLOW
LOW BATTERY
delivered
AC POWER
LOW BATTERY
LOW BATTERY
MALFUNCTION
MALFUNCTION

Turning off the pump

MALFUNCTION
AC POWER
AC POWER
AC POWER
800
ENTERAL FEEDING PUMP
RATE ML/HR
RUN
HOLD
DOSE COMPLETE
VOL
OCCLUSION
DOSE
FED
NO FLOW
LOW BATTERY
DEC
INC
MALFUNCTION
AC POWER
1
HOLD
DOSE
DEC
RUN
HOLD
HOLD
> Turn on the pump by pressing the "POWER/ON" key. The pump is in "HOLD" mode.
VOL
DOSE
DOSE
FED
> Set the desired flow rate (ml/h) by pressing the "INC" or "DEC" keys.
DEC
DEC
INC
> Flow rates can be set from 1 to 300 ml/h, in 1 ml increments.
> Press the "RUN" key to start the pump.
RUN
HOLD
RUN
HOLD
HOLD
RUN
HOLD
VOL
> After programming the flow rate do not immediately press the "RUN" key.
DOSE
VOL
FED
DOSE
DOSE
FED
> Press the "DOSE" key to set the desired dose.
VOL
DOSE
FED
DEC
> Press "INC" or "DEC" to program the correct dose (in ml).
INC
DEC
DEC
INC
DEC
> Press the "RUN" key to start the pump.
INC
RUN
HOLD
> Press "HOLD" key, pump will stop, all memory and programming is retained.
VOL
RUN
HOLD
DOSE
FED
Use this to: change flow rate, alter dose setting, change feed, administer medication.
RUN
HOLD
VOL
DEC
> Press "RUN" to restart the pump.
DOSE
INC
FED
RUN
RUN
HOLD
HOLD
VOL
DOSE
FED
DEC
VOL
> Press "VOL FED" key, display will indicate total volume delivered.
INC
VOL
DOSE
DOSE
FED
FED
DEC
INC
DEC
DEC
INC
INC
> Turn on the pump by pressing the "OFF" key for ca 3 seconds.
ENTERAL FEEDING PUMP
RATE ML/HR
DOSE COMPLETE
OCCLUSION
NO FLOW
LOW BATTERY
MALFUNCTION
AC POWER
800
ENTERAL FEEDING PUMP
RATE ML/HR
RUN
HOLD
DOSE COMPLETE
VOL
OCCLUSION
DOSE
FED
NO FLOW
LOW BATTERY
DEC
INC
MALFUNCTION
AC POWER
2
RUN
HOLD
VOL
DOSE
FED
DEC
INC
800
ENTERAL FEEDING PUMP
RATE ML/HR
RUN
HOLD
DOSE COMPLETE
VOL
OCCLUSION
DOSE
FED
NO FLOW
LOW BATTERY
DEC
INC
MALFUNCTION
AC POWER
3
4

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Flocare 800 and is the answer not in the manual?

Questions and answers

Summary of Contents for Nutricia Flocare 800

  • Page 1: Operating Instructions

    Flocare 800 Short Manual ENTERAL FEEDING PUMP RATE ML/HR HOLD DOSE COMPLETE OCCLUSION DOSE NO FLOW LOW BATTERY MALFUNCTION AC POWER Preparation and start of the pump ENTERAL FEEDING PUMP ENTERAL FEEDING PUMP ENTERAL FEEDING PUMP ENTERAL FEEDING PUMP RATE ML/HR...
  • Page 2 Flocare 800 Problem solving instructions Error messages and instructions Please refer to detailed Instructions for Use for additional information Problem/Alarm Likely cause Instructions > The pump has been left in “hold” mode > Reset the alarm by pressing the “HOLD” key, or HOLD for 3.5 minutes.