SIKA TP17 165 Operating Manual

SIKA TP17 165 Operating Manual

Temperature calibrators tp17 series; tpm series
Table of Contents

Advertisement

Operating manual
Operating manual .................................................... page 1 - 60
Temperature calibrators
Series TP17 / TPM
Please keep this operating manual for future reference.
If the device is resold, please provide the operating manual along with it.
(Translation)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TP17 165 and is the answer not in the manual?

Questions and answers

Summary of Contents for SIKA TP17 165

  • Page 1 Operating manual (Translation) Operating manual ............ page 1 - 60 Temperature calibrators Series TP17 / TPM Please keep this operating manual for future reference. If the device is resold, please provide the operating manual along with it.
  • Page 2: Table Of Contents

    Main menu ......................... 30 5.4.1 Automatic control ....................32 5.4.2 Manual control ...................... 33 5.4.3 Set point memory (S version) ................34 5.4.4 Gradient control / temperature profile (S version) ..........35 - 2 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 3 TP17 / TPM Testing process / Calibration ..................41 Testing temperature sensors ................... 41 Testing process ......................41 Problems ........................43 Return shipment to the manufacturer ..............44 Maintenance, cleaning and transport ................. 45 Maintenance ......................45 8.1.1 Recalibration ......................46 Cleaning ........................
  • Page 4: About This Operating Manual

     Check the specified points or notices. information that is marked with this  Reference to another section, document or symbol. source.  Follow the specified instructions and steps. • Item. Adhere to the given order. - 4 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 5: Device Description

    TP17 / TPM Device description Device description The calibrators of the series TP17 / TPM serve for testing and calibrating different temperature measuring instruments and temperature sensors, as well as for measurement of temperatures. The portable instruments are of compact and robust construction and thus allow use directly on-site or in a laboratory.
  • Page 6: Delivery, Unpacking And Accessories

     Use the type plate to check if the delivered unit corresponds to your order.  In particular, for devices with electrical components, check to see if the correct power supply voltage is specified. - 6 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 7: Intended Use

    The calibrator is under guarantee for 12 months as from the date of delivery for construction errors or material defects. The guarantee is limited to repair or replacing the calibrator. SIKA also provides an extra 5-year guarantee for calibrators which are calibrated and inspected annually by the SIKA-DAkkS laboratory.
  • Page 8: Exclusion Of Liability

    In order to guarantee that the device operates safely, the operator must act competently and be conscious of safety issues. SIKA provides support for the use of its products either personally or via relevant literature. The customer verifies that our product is fit for purpose based on our technical information.
  • Page 9 The calibrator cannot then be started any more.  After it has cooled, send the calibrator for inspection to SIKA. Risk of injury from hazardous gases! When liquids are heated, the evaporation can result in dangerous gases being formed.
  • Page 10: Safety Instructions For The Application Of Calibration Liquids

    Mineral oil: SIKA supplies the calibrators only with silicone oil. Using mineral oil is possible, but must be done on your own responsibility. The danger and the risk must be borne by the customer and our warranty will be rendered null and void.
  • Page 11: Construction And Function

    TP17 / TPM Construction and function Construction and function 3.1 Construction  The calibrator consists of a robust, black-and-red painted steel housing and has a  carrying handle at the top  The front part of the housing holds the electronics of the TP17 / TPM, the controller with ...
  • Page 12: Functions

    The 5-pole socket is provided with two connections, A and B, which have to be connected to the respective sockets of the PC, the level converter or the network. 5-pole socket - 12 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 13 Certain factory settings have to be carried out to enable configuration of a network. In this case, please contact your supplier or SIKA directly. Network configuration PLEASE NOTE: If you access the programming via the keypad while communication via a serial interface is running, the message "buSy"...
  • Page 14: Commissioning And Operation

    The plug of the mains connecting cable serves as a "KILL" switch.  Ensure that the plug is always easily accessible and easy to reach.  In an emergency, pull the plug, so that the machine is isolated from the mains. - 14 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 15: Electrical Connection

    IMPORTANT! Mains connecting cable! The mains connecting cable may only be replaced by an equivalent cable.  Use only original cables from SIKA or approved cables of the same type with the correct design as replacements ( § 10.1…10.4 "Electrical characteristics").
  • Page 16: Preparation Of The Calibrator

    It is thus easily possible to switch between the dry block, infrared, surface and micro bath functions. Which measuring insert you can use in which calibrator type is shown in the table in § 3.2. - 16 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 17: Dry Block Calibrator

    TP17 / TPM Commissioning and operation 4.3.1 Dry block calibrator Type TP17 200 TP17 165 TP17 166 TP17 450 TP17 650 TPM 165 TPM 225 TPM 255 Dry block         Adapter sleeves with single or multiple holes are used for the calibration of straight temperature sensors.
  • Page 18: Calibrator For Infrared Thermometer

    Commissioning and operation TP17 / TPM 4.3.2 Calibrator for infrared thermometer Type TP17 200 TP17 165 TP17 166 TP17 450 TP17 650 TPM 165 TPM 225 TPM 255 Infrared      A special infrared insert is used for the contact-less measurement infrared thermometers.
  • Page 19: Calibrator For Surface Temperature Sensors

    TP17 / TPM Commissioning and operation 4.3.3 Calibrator for surface temperature sensors Type TP17 200 TP17 165 TP17 166 TP17 450 TP17 650 TPM 165 TPM 225 TPM 255 Surface      A special surface insert is used for calibrating surface temperature sensors. This makes calibration fast and easy.
  • Page 20: Micro Bath Calibrator

    Commissioning and operation TP17 / TPM 4.3.4 Micro bath calibrator Type TP17 200 TP17 165 TP17 166 TP17 450 TP17 650 TPM 165 TPM 225 TPM 255 Micro bath    CAUTION! Risk of injury - Wear safety goggles! Calibration liquid can be ejected when working with the micro bath calibrator.
  • Page 21 TP17 / TPM Commissioning and operation Operation of magnetic stirrer: The speed of the magnetic stirrer is set via the thumb wheel ( Fig.) next to the controller. Set the speed to the respective maximum to achieve a thorough mixing. ...
  • Page 22 Rise due to stirring The rotation of the magnetic stirrer forms a whirlpool in the liquid. This raises the fill level at the wall. max. fill level of the micro bath - 22 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 23 TP17 / TPM Commissioning and operation Tank: The maximum fill level in the tank is displayed by the upper edge of the aluminium lining. The maximum fill level is ~0.45 litres. Tub insert: The maximum fill level with the tub insert is below the fixture for the sleeve exchange tool. The maximum fill level is ~0.32 litres.
  • Page 24: Switching On, Cool Down And Switching Off

    2 indicates the status of the output for the active cooling. If the LED OUT 2 lights up, the active cooling is running. If the LED OUT 2 does not light up, the cooling is not active. - 24 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 25: Operating Modes

    TP17 / TPM Commissioning and operation IMPORTANT! Mains failure or separation from the mains! If there is a mains failure, or if the main switch is turned off, or upon removal of the mains connection ("EMERGENCY STOP"), the built-in fan does not blow any more cooling air.
  • Page 26: Operation

    • Reducing values to be set.  If LED OUT 2 is not lit, the calibrator or micro • Selection of individual menu items. bath is not cooling. • Return to previous menu level. - 26 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 27: Calibration (Calibration Mode)

    TP17 / TPM Operation 5.2 Calibration (calibration mode) As soon as the calibrator has been switched on, it switches to the calibration mode after initialization 5.2.1 Display of reference and set temperature The current reference temperature is displayed in the upper and the set temperature in the lower of the display.
  • Page 28: Display Of Fan/ Cooling Control

     The current number of operating hours of the calibrator is displayed for ~5 s. 5.2.5 ConF menu In the ConF menu, system parameters of the calibrator can be set. These settings may only be carried out by authorised SIKA personnel. This menu is protected by a password. CAUTION! Material damage! The ConF menu is protected by a password.
  • Page 29: Set A Temporary Set Temperature (Set Point Mode)

    TP17 / TPM Operation 5.3 Set a temporary set temperature (set point mode) In this operating mode it is possible to temporarily modify a saved set temperature.  Press the P key shortly. The currently active set point memory, e.g. SP 2 (set point 2), is displayed by the upper display.
  • Page 30: Main Menu

    For this, four additional options are available in the main menu: Micro bath function. Dry block function. Infrared black body function. Surface temperature calibration function. - 30 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 31 TP17 / TPM Operation In calibration mode, the upper display will then show the chosen linearisation (LI, DB, Ir or SU) every 5 seconds in alternation with the actual temperature. The internal references are set to normal when operating the micro-bath, the dry block and the infrared function;...
  • Page 32: Automatic Control

    P key. PLEASE NOTE: The control has been reactivated. The calibrator / micro bath is in calibration mode and the set temperature is targeted. - 32 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 33: Manual Control

    TP17 / TPM Operation 5.4.2 Manual control Switching on the manual control: It is possible to switch off the automatic control of the calibrator / micro bath and to achieve the desired temperature via manual control.  Press the P key for approx. 5 sec., the main menu opens. The last selected function appears on the upper display.
  • Page 34: Set Point Memory (S Version)

     Press the P key to confirm the set temperature. The set point memory closes and the display returns to the parameter level. key to return to the calibration mode.  Press and hold the   - 34 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 35: Gradient Control / Temperature Profile (S Version)

    TP17 / TPM Operation If no key is pressed for approx. 15 seconds, the device automatically returns to a previous level up to the calibration mode. Retrieving the saved set temperatures (S version): The saved set temperatures can be retrieved in calibration mode. ...
  • Page 36 Duration time „dur.t“: The duration time "dur.t" is active if the set temperature SP1 has been achieved. Subsequently, the calibrator / micro bath automatically switches to set temperature SP2. - 36 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 37 TP17 / TPM Operation ACTIVATE TEMPERATURE PROFILE! If you have carried out settings for these three settings, the calibrator / micro bath uses the new values only when modifying the set temperature or switching the calibrator off and on again. A further procedure is to switch off the automatic control prior to modifying parameters and to switch it on again afterwards (...
  • Page 38 Activate temperature profile! After carrying out the settings, the calibrator uses the new values only when modifying the set temperature or switching the calibrator / micro bath off and on again. - 38 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 39 TP17 / TPM Operation 5.4.4.2 Setting the cooling gradient (S version) The cooling gradient "SLoF" is active if the reference temperature is higher than the set temperature. The setting range extends from 99.99 °C/min up to 0.00 °C/min. PLEASE NOTE: The function is deactivated if SLoF = InF (In no Function) has been set.
  • Page 40 After carrying out the settings, the calibrator / micro bath uses the new values only when modifying the set temperature or switching the calibrator / micro bath off and on again. - 40 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 41: Testing Process / Calibration

    TP17 / TPM Testing process / Calibration Testing process / Calibration CAUTION! Risk of burns! The calibrator can become very hot when in operation. Touching hot parts can result in serious injuries.  Never touch the metal block, the tank, the adapter sleeve or the test specimen at temperatures above 35 °C or below 10 °C.
  • Page 42  Note the instructions in the section "Switching on, cool down and switching off" ( § 4.4).  If necessary drain the tank or tub insert of the micro bath calibrator with the drain pump.  Clean the calibrator ( § 8.2). - 42 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 43: Problems

    If you are unable to remedy any particular problem, then immediately disconnect the calibrator in order to protect it from unintended operation. Contact your supplier or directly to SIKA. Please send the device for repair with a brief description of the problem, the environmental conditions and the length of time the device was operational before the problem occurred.
  • Page 44: Return Shipment To The Manufacturer

    Due to legal requirements placed on environmental protection and occupational safety and health and to maintain the health and safety of our employees, all units returned to SIKA for repair must be free of toxins and hazardous substances. That also applies to cavities in the devices.
  • Page 45: Maintenance, Cleaning And Transport

    Recalibration:  Send the calibrator for Recalibration ( § 8.1.1) once annually or after a maximum of 500 operating hours to SIKA. Calibration liquid: Calibration liquids get fouled or age with time. This depends to a great extent on the type of liquid and the usage behaviour.
  • Page 46: Recalibration

    The calibrator should, depending on the application situation, be inspected at appropriate intervals on the basis of DIN ISO 10012. We recommend you to return the calibrator to SIKA at intervals of max. 12 months or approx. 500 operating hours for recalibration and readjustment.
  • Page 47: Cleaning

    IMPORTANT! Temperature fuse at over-temperature! An air flow that is too low can result in the temperature fuse getting triggered. The calibrator is then rendered operationally disabled, and must be sent to SIKA.  Ensure that the grille openings are always clear.
  • Page 48: Transport

    Let the calibrator neither fall nor tip over.  Protect the calibrator from inadequate environmental influences. Use our transport case which is available as accessories. Carry the calibrator only at the handle.  - 48 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 49: Decommissioning And Disposal

    The calibrator of the series TP17 / TPM consist of various different materials. It must not be disposed of with household waste (EU Directive 2002/96/EC).  Take the TP17 / TPM to your local recycling plant  send the TP17 / TPM back to your supplier or to SIKA. Technical changes reserved - 49 -...
  • Page 50: Technical Data

    0…50 °C Transport and storage -10...60 °C temperature Humidity (RH) 30…95 % (non-condensing environment) Operating conditions: - Location • Altitude Interiors • up to 2000 m - Operating position Standing upright/vertically - 50 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 51: Characteristics Series Tp 17

    TP17 / TPM Technical data 10.2 Characteristics series TP 17 … Series Characteristics 17 165 17 166 17 200 17 450 17 650 Calibrator Temperature range -35…165 °C -35…165 °C -55…200 °C T …450 °C …650 °C Accuracy ± 0.4 °C ±...
  • Page 52: Characteristics Tp 17

    300 mm 270 mm Weight ~10.0 kg ~10.0 kg ~12.5 kg ~7.5 kg Sleeves for smaller ∅ 1.5…25 mm 1.5…55 mm 1.5…25 mm 1.5…55 mm 1.5…25 mm (in 0.5 mm steps) - 52 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 53: Characteristics Series Tp M

    TP17 / TPM Technical data 10.4 Characteristics series TP M … S Series TP M 165 S TP M 225 S TP M 255 S Characteristics Calibrator Temperature range: - Dry block -35…165 °C …225 °C …255 °C - Infrared -35…165 °C …225 °C …255 °C...
  • Page 54: Heating And Cooling Periods

    Reference sensor Pt100, Ø 6 mm, stainless steel casing. • The reference sensor is located 5 mm above the mesh insert in the middle of the tank. 10.5.1 TP 17 165; TP 17 166, TP 17 200 (dry block) - 54 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 55: Tp 17 450; Tp 17 650 (Dry Block)

    TP17 / TPM Technical data 10.5.2 TP 17 450; TP 17 650 (dry block) 10.5.3 TP 17 166 S; TP 17 450 S (infrared, surface) Technical changes reserved - 55 -...
  • Page 56: Tp M 165 S

    Technical data TP17 / TPM 10.5.4 TP M 165 S Dry block, infrared and surface Water and silicone oil - 56 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 57: Tp M 225 S, Tp M 255 S

    TP17 / TPM Technical data 10.5.5 TP M 225 S, TP M 255 S Dry block, infrared and surface Water and silicone oil Technical changes reserved - 57 -...
  • Page 58: Guideline Dakks-Dkd-R 5-4

    If control measurements with a calibrated thermometer are not taken annually, recalibration of the temperature block calibrator is strongly recommended. © Deutsche Akkreditierungsstelle GmbH (German Accreditation Body) • E-Mail: contact@dakks.de • Internet: www.dakks.de - 58 - © SIKA • Ba_TP17-TPM_en • 01/2016...
  • Page 59 TP17 / TPM Guideline DAkkS-DKD-R 5-4 • Appendix B For your notes Technical changes reserved - 59 -...
  • Page 60 TP17 / TPM Sensors and Measuring Instruments Flow Measuring Instruments Test and Calibration Instruments SIKA Dr. Siebert & Kühn GmbH & Co. KG Struthweg 7–9 D-34260 Kaufungen  Germany  +49 (0)5605 803-0  +49 (0)5605 803-54 info@sika.net  www.sika.net - 60 - ©...