Information - Neurotech AvivaTens XP Instructions For Use & Warranty

Pain management system
Table of Contents

Advertisement

PROGRAM INFORMATION
Program
Program 1 (High Frequency Trigger):
Program 1 delivers a stimulus of 4Hz. When the Trigger button is pressed a high frequency 99Hz stimulus is
delivered. To disable High Frequency Trigger mode press the Trigger button a second time.
Program 2, 3 (Burst Mode):
Burst mode is available on Programs 2 and 3. When the Burst button is pressed Burst mode is enabled. To
disable Burst mode press the button a second time. If Burst mode is selected during Program 2, it results in a
frequency of 99Hz at a pulse width of 120 s for 3 seconds. This then returns to the normal program frequency and
pulse width for 1.5 seconds. If Burst mode is selected during Program 3, it results in a frequency of 4 Hz at
a pulse width of 120 s for 3 seconds, before returning to the normal program frequency and pulse width
23
for 3 seconds.
Frequency
Pulse duration
No.
(Hz)
(µsec)
1
4
150
2
35
150
3
99
150
4
Channel 1: 4
150
Channel 2: 99
5
4 - 99
150
Treatment
Time (min)
Burst or Trigger
open
High Frequency Trigger
open
Burst – Frequency of 99Hz, pulse
width 120 µsec. for 3 seconds;
Will return to program settings
for 1.5 seconds.
open
Burst – Frequency of 4Hz, pulse
width 120 µsec. for 3 seconds;
Will return to program settings
for 3 seconds.