Optimum Battery Operation; Rechargeable Battery Capacity; Replacing The Sensor And Sensor Filter - BW Technologies CONNEX1 Operator's Manual

Single gas detector
Hide thumbs Also See for CONNEX1:
Table of Contents

Advertisement

Maintenance

16.1.1 Optimum Battery Operation

To ensure maximum use of the battery, perform the following:
• To obtain full operating capacity, allow the battery to fully charge and
discharge.
• To achieve the maximum number of charges, ensure the battery is
charged between 5°C and 40°C (41°F and 104°F). Do not charge
the battery in temperatures above 40°C (104°F) or below 5°C (32°F).
If the battery is charged outside of the charging temperature
specifications, a charging error will result.

16.1.2 Rechargeable Battery Capacity

A rechargeable battery's runtime decreases approximately 20% over a two-year
period of typical use.
To maximize the battery's capacity over its lifetime, charge the battery at 20°C.

16.2 Replacing the Sensor and Sensor Filter

WARNING
To avoid personal injury and/or property damage, use only
sensors that are specifically designed for the detector. Refer to
Replacement Parts and Accessories on page
36
19.
53.
CAUTION
The sensor has a high degree of resistance to common vapours and gases.
To clear a sensor, move the detector to a clean environment and wait 10 to
30 minutes.
To prevent accidental poisoning of the sensor, refer to
Contaminants on page
5.
1.
Press and hold
to deactivate the detector.
2.
Remove the four machine screws from the rear shell.
1.3 Sensor Poisons and
ConneX1 Operator's Manual