Parameter
6490-9000 Revision 3
Buzzer Mute Time
Sets a time (0-59 minutes) during which the active
buzzer remains muted:
after the "I" button is pressed, or
after Modbus register 4000 is set to 0.
Baud Rate
Sets the baud rate for Modbus (RS-485)
communications.
00
9,600 baud (default)
01
19,200 baud
Stop Bits
Sets the number of stop bits required to match the
controlling
communications
building management system, etc.).
01
1 stop bit (default)
02
2 stop bits
Parity
Sets Modbus parity option.
00
None (default)
01
Odd parity
02
Even parity
Description
equipment
MGS-250 Manual
(e.g.,
37