Download Print this page

AEMC SR751 User Manual page 4

Ac current probe

Advertisement

OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model SR751/SR752
Connect the black lead of the current probe to "Common" and the red lead to the AC voltage input on
your DMM or other voltage-measuring instrument. The AC current probe has a ratio of 1000:1. This
means that for 1000A AC in a conductor around which the probe is clamped, 1V AC will come out of
the probe leads to your DMM or instrument. The output is 1mV AC/A AC. Select the range which
corresponds to the measured current. If the current magnitude is unknown, start with the highest range
and work down until the appropriate range and resolution are reached. Clamp the probe around the
conductor. Take the reading on the meter and multiply it by 1000 to obtain the measured current (e.g,
260mV reading = 260 x 1000 = 260,000mA or 260A)
For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE
Warning:
For maintenance use only original replacement parts.
To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into
the probe
Cleaning:
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine 600)
to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will ensure
that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the
outside of the shipping container.
Chauvin Arnoux
15 Faraday Drive  Dover, NH 03820 USA
(800) 945-2362 (Ext. 360) • (603) 749-6434 (Ext. 360) • repair@aemc.com
(Or contact your authorized distributor)
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or application
of this instrument, please contact our technical hotline:
(800) 343-1391 • (508) 698-2115 • techsupport@aemc.com
LIMITED WARRANTY
The current probe is warranted to the owner for a period of two years from the date of original purchase
against defects in manufacture. This limited warranty is given by AEMC
from whom it was purchased. This warranty is void if the unit has been tampered with, abused or if the
defect is related to service not performed by AEMC
Full warranty coverage and product registration is available on our website at:
Please print the online Warranty Coverage Information for your records.
®
®
, Inc. d.b.a. AEMC
Instruments
®
Instruments.
®
Instruments, not by the distributor
www.aemc.com/warranty.html
99-MAN 100167 v14 05/17

Advertisement

loading

This manual is also suitable for:

Sr752