Procedures (Ohms Output); Procedures (Ma Input); Procedures (Idos Umm) - Test Equipment Depot Druck DPI 811 User Manual

Rtd calibrator and rtd loop calibrator
Table of Contents

Advertisement

3. To make sure that the calibration is correct, select the
applicable ohms input task (Table 2).
4. Make a 4 wire connection to the applicable standard
resistor (Table 10) and measure the value (Figure 1).
5. Make sure that the error is in the specified limits
(Table 10).
Table 10: Ohms input error limits
Standard
Resistor
Resistor*
error
(Ω)
(Ω)
0 (Short circuit)
-
100
0.008
200
0.013
300
0.018
400
0.007
1k
0.042
2k
0.052
4k
0.072
*
Or an equivalent resistance simulator
Calibration - Procedures (Ohms output)
1. Connect the instrument to the calibration equipment
(Figure 2).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a two-point
calibration for each range.
• Range: 0-399.9Ω
• Range: 400Ω-1999.9Ω
• Range: 2kΩ-4kΩ
The display shows the applicable instructions to
calibrate each range.
4. To make sure that the calibration is correct, select the
applicable ohms output task (Table 2).
5. Supply the specified values (Table 11). Make sure that
the error is in the specified limits.
K345 Issue 4
Permitted
DPI 811/812 error
(Ω)
0.05
0.05
0.05
0.05
0.05
0.25
0.25
0.50
Table 11: Ohms output error limits
Ohms
Calibrator
(Ω)
error
(Ω)
0
0.003
100
0.004
200
0.005
300
0.007
400
0.008
1000
0.015
2000
0.026
4000
0.049
Calibration - Procedures (mA input)
1. DPI 812 only. Connect the instrument to the calibration
equipment (Figure 5).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a three-point
calibration (-FS, Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable mA input task (Table 2) and apply these
values:
• mA: -55, -40, -24, -18, -12, -6, 0 (open circuit)
Then mA: 0, 6, 12, 18, 24, 40, 55.
5. Make sure that the error is in the specified limits
(Table 12).
Table 12: mA input error limits
Applied
Calibrator
mA
error
(mA)
±55
0.0022
±40
0.0018
±24
0.0014
±18
0.0004
±12
0.0003
±6
0.0002
0 (open circuit)
-
Calibration - Procedures (IDOS UMM)
Refer to the user manual for the IDOS UMM.
When the calibration is complete, the instrument
automatically sets a new calibration date in the UMM.
Permitted
DPI 811/812 error
(Ω)
0.05
0.06
0.06
0.07
0.07
0.30
0.40
0.80
Permitted
DPI 811/812 error
(mA)
0.005
0.004
0.003
0.003
0.002
0.002
0.001
11