Sample Measurement; Ph Measurement; Mv Measurement; Temperature Measurement - OHAUS ST300 Instruction Manual

Portable ph meter
Hide thumbs Also See for ST300:
Table of Contents

Advertisement

EN-12
4.2

Sample measurement

4.2.1

pH measurement

Place the electrode in the sample, stir for 10 seconds, and wait 30-60 seconds.
Press button-
during measurement.
When meter reaches the proper value, the pH value with the temperature is displayed on the
screen.
4.2.2

mV measurement

Press button-
Follow the same procedure as for pH measurement to perform a mV measurement.
4.3

Temperature measurement

For better accuracy, we recommend to use either a built-in or a separate temperature electrode.
If a temperature electrode is used, ATC and the sample temperature are displayed.
If the meter does not detect a temperature electrode, it automatically switches to the manual
temperature compensation mode and MTC appears. MTC temperature should be set. (See
section 3.1 in this manual)
Note: STARTER 300 accepts NTC 30 kΩ temperature sensor.
4.4

Using the memory

4.4.1

Storing a reading

The STARTER 300 can store up to 30 results.
Press button-
result has been stored.
If you press button-
store further data you will have to clear the memory. (See 4.4.3 below)
4.4.2

Recalling from memory

Press and hold button-
measurement reaches its final reading and freezing.
Press button-
which result is being displayed.
Press button-
to start the pH measurement,
to switch between pH and mV measurement.
when the measurement reaches its final reading, M01 indicates that one
when M30 is displayed, FUL displays to indicate the memory is full. To
to recall the stored values from memory when the current
or button-
to scroll through the stored results. R01 to R30 indicates
to exit.
STARTER 300 Portable pH Meter
appears on the display.
is blinking