Section 1: Description And Specifications; Features And Applications - Emerson Rosemount 396PVP Instruction Manual

Ph/orp sensors
Hide thumbs Also See for Rosemount 396PVP:
Table of Contents

Advertisement

Instruction Manual
LIQ-MAN-396P-396PVP

Section 1: Description and Specifications

1.1
Specifications - General
Measurements and Ranges: pH*: 0-14 / ORP: -1500 to 1500 mv
Available pH ACCUGLASS Types: GPLR hemi or flat glass
Wetted Materials: Titanium, Polypropylene, EPDM, glass; platinum (ORP only)
Process Connection: 1 in. MNPT front and rear facing threads
Temperature Range: 0-100°C (32-212°F)
Pressure Range-Hemi bulb: 100-1135 kPa [abs] (0-150 psig)
Pressure Range-Flat bulb: 100-790 kPa [abs] (0-100 psig)
Minimum Conductivity: 100 µS/cm
Integral Cable 396P: Code 01 - 25 ft; Code 02 - 15 ft coaxial / 396PVP: none - must use mating
VP cable
Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/2 lb)
*Percent Linearity
pH Range
GPHT Hemi
0-2 pH
94%
2-12 pH
99%
12-13 pH
97%
13-14 pH
92%
Description and Specifications
396 / 396VP
GPHT Flat
95%
99%
96%
Section 1: Description and Specifications
396P / 396PVP
GPHT Hemi
GPLR Flat
94%
97%
98%
98%
95%
98%
March 2017
396R / 396RVP
GPHT Hemi
GPHT Flat
94%
93%
97%
98%
98%
95%
98%
1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 396p

Table of Contents