Figure 2-5 Jet Spray Cleaner PN 12707-00
Note: Sensor must be installed at least
10° above the horizon)
Figure 2-6 Low Flow Panel PN SQP10077-LQD
Installation
The jet spray cleaner eliminates routine, manual sensor
maintenance by cleaning the sensor with water or
compressed air. Flow through the cleaner can be
controlled by a solenoid valve.
Note: The jet spray cleaner can be used with
handrail mounting assembly (PN 11275-01, not
shown) or can be mounted through conduit as shown
below.
Jet Spray Cleaner with pH sensor
Low Flow Panel Specification
Inlet flow
Inlet pressure
Temperature
*The minimum inlet pressure is required to open a check valve, which
prevents the flow cell from draining if sample flow is lost. Removing
the check valve lowers the inlet pressure requirement to a few feet of
water head.
Section 2: Installation
April 2017
3-80 gph (11.4 – 304 L/hr)
3-65 psig (122 – 549 kPa abs)*
32 to 122 °F (0 to 50 °C)