5.9.2 Product Regulator Cleaning or Rebuilding

Clean or rebuild the product regulator if the regulator cannot be adjusted for
lockout.
1. Set the unit's
2. Remove the right side panels and back panels.
3. Use large pliers to unscrew the bonnet of the product regulator, which
4. Remove the diaphragm. (Clean or replace it.)
5. Using a Phillips-head screwdriver, unscrew the diaphragm stem guide
6. Remove the seat. Be careful not to lose the spring located behind the seat.
7. Clean or replace the seat.
8. With the spring behind the seat, screw the diaphragm stem guide back into
9. Install a clean or replacement diaphragm.
10. Put the large spring and slip ring into the bonnet, and screw the bonnet onto
11. Reset the product regulator as described in Section 5.9.1 .

5.10 Circuit Breaker Replacement

Circuit Breaker Removal
To remove the circuit breaker, follow the steps below:
1. Set the unit's
2. Remove the left side panel.
3. Disconnect the circuit breaker leads.
4. Unscrew the circuit breaker while you apply pressure to the circuit breaker
Circuit Breaker Installation
Follow the removal procedure for the circuit breaker in reverse order to install the
new circuit breaker. Wire connections on circuit breaker can be made to either
terminal.
5.11
/0 Power Switch Replacement
I
I
/0 Power Switch Removal
To remove the
1. Set the unit's
2. Remove both side panels and back panel.
3. Disconnect wire connections from the
4. Push on the back of the power switch, while holding in its upper and lower
/0 switch to the "0" position and disconnect the power cord.
I
contains a large spring.
Adjust the product regulator fully counter-clockwise to unload the
spring. This makes disassembly and reassembly easier.
located in the center of the regulator body to gain access to the seat.
the body of the regulator. (Do not over tighten.)
the regulator body.
/0 switch to the "0" position and disconnect the power cord.
I
retaining ring.
I
/0 power switch, follow the steps below:
/0 switch to the "0" position and disconnect the power cord.
I
retaining tabs, and remove the switch through the front panel.
I
/0 power switch.
17
MN134-1 rev C