Start Up/Battery Test; Power Failure Alarm Test; Oxygen Concentration Test And Specification - AirSep NewLife Intensity 10 Service Manual

Oxygen concentrator
Hide thumbs Also See for NewLife Intensity 10:
Table of Contents

Advertisement

oxygen monitor detects lower than therapeutic levels of oxygen concentration. The alarm
remains on until you correct the alarm condition or you set the
position. Refer to Section 6.0 for a list of probable alarm causes.

2.3.1 Start Up/Battery Test

Each time the NewLife Intensity 10 unit is turned on, an alarm should sound for
approximately five seconds.

2.3.2 Power Failure Alarm Test

To test the power failure alarm, disconnect the unit from power, set the
switch to the "
to the troubleshooting chart in Section 6.3 of this manual.

2.4 Oxygen Concentration Test and Specification

To ensure that the unit's output of oxygen is within specification, you must perform a test
of the oxygen concentration. Test the unit upon delivery to a patient and at periodic
intervals. Equipment Providers need to establish and implement a protocol to check
oxygen concentration.
1.
If a humidifier bottle is used, disconnect it.
2.
Connect a calibrated oxygen concentration analyzer to the oxygen outlet.
Verify that the product flow rate delivered by the unit matches the patient's
3.
prescription and does not exceed the capacity of the unit.
Set the unit's
4.
minutes for the oxygen concentration to stabilize. Take oxygen concentration
readings and verify levels are within specification at the liter flow being tested.
5.
Disconnect the oxygen analyzer, and reconnect the humidifier bottle (if used).
The audio alarm must sound for approximately five seconds each time
the unit is turned to the "
condition.
" position. This should activate the audio alarm. If it does not, refer
I
/0 power switch to the "
l
Do not measure oxygen concentration output after the product stream
passes through a humidifier bottle, or erroneous readings will result.
AirSep NewLife Intensity 10 Concentration Specifications
Liter
Flow
Oxygen Concentration
2-9 lpm
92% +3.5/-3%
10 lpm
90% +5.5/-3%
" position to indicate the battery is in good
I
" position. Allow approximately five
I
6
/0 power switch to the "0"
I
/0 power
l
In Spec
89% or higher
87% or higher
MN134-1 rev C