Chapter 2
The communications cable from the optical bench or from another dispersion unit
connects here.
If more than one dispersion unit is connected to the system and this is the first dis-
persion unit in line, a communication cable will be connected from this connector
to the Accessory comms - in connector on the second dispersion unit.
The manual drain button can be pressed to drain the well. Users may wish to do
this if, for example, they need to drain the dispersion unit without having to start
the control software. Pressing the button once (for about half a second) will open
the drain valve. Pressing the button a second time will close the drain valve. If the
drain valve is left open for more than 15 minutes it will automatically close.
Page 2-8
Warning!
The PSU is not designed for operation in wet conditions. Do not place it
in areas prone to spillages.
Warning!
Only use the PSU and cables supplied by Malvern Instruments.
Accessory comms "in" connector
Accessory comms "out" connector
Caution!
The termination plug supplied with the optical bench must be fitted here
if this is the only dispersion unit connected or it is the last dispersion unit
in the line.
Manual drain
Hardware features
MAN 0389