2.03 Setting Generator Level

GENERATE
Modulation
"L" Icon will flash. Enter the Level using the keypad, then press your desired units. The
COM3010 is now generating your desired level.
Example:
To cancel an entry, press
Property of Ramsey Electronics, Inc.
Do not reproduce or distribute.

2.04 Setting Generator Steps/Increments

GENERATE
Modulation
setting make sure the level field is selected (blinking icon) or press LEVEL, and use the
and
DOWN
arrow keys.
Level can be entered in dBm or volts and can be stepped in dBm or volts. dBm can be
stepped in as little as 0.1 dB, and volts down to 0.1uV. Remember that the basis for dB is
logarithmic and volts are linear so any steps entered in volts when viewing dBm will not di-
rectly correlate, likewise with steps entered in dBm when volts are displayed.
It is simple to determine how many volts there are in the current setting. While still in the
Level field, press
kHz/mV
press MHz/dBm.
Example:
TX Level
Seq/Reg
TX Level
Seq/Reg
or Hz/uV. The display should now show 652.4uV. To switch back,
15
BASIC OPERATION
RF Level can be entered in either the dBm or
Volts scale. The Range is 0 dBm (223 mV) to
-140 dBm (0.022uV) in 0.1 dB steps.
To enter the Level Press the
To enter a new level increment, press
then
LEVEL
and the current increment value
will be displayed. To enter a step size of
0.1dB, enter
0.1
then -dBm/MHz. To cancel
out press
SHIFT
then ENTER. To step the level
LEVEL
key. The
SHIFT
UP