6400
Calorimeter Control Limit Values in BTU/lb When Benzoic Acid is Used as a Test Sample
Accepted heat of combustion taken as 11373 BTU/lb.
Instrument precision 0.10%.
Control limits based on 99% confidence (3 sigma) values.
Values are in BTU/lb.
NUMBER OF
UCL FOR THE RANGE
OBSERVATIONS
(HIGH – LOW) WITHIN
IN A GROUP
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
Table 6-3
THE GROUP
41.9
49.6
53.4
55.9
57.8
59.2
60.4
61.3
62.2
62.9
63.6
64.2
64.8
65.3
65.8
66.2
66.6
67.0
67.4
67.7
68.0
68.3
68.6
68.9
UCL FOR THE
DEVIATION OF THE GROUP MEAN
RSD WITHIN
FROM THE ACCEPTED VALUE OR
THE GROUP
0.261%
0.228%
0.209%
0.196%
0.187%
0.181%
0.175%
0.171%
0.167%
0.164%
0.161%
0.159%
0.156%
0.154%
0.153%
0.151%
0.150%
0.148%
0.147%
0.146%
0.145%
0.144%
0.143%
0.142%
Standardizations
MAXIMUM PERMISSIBLE
GRAND MEAN
34.1
24.1
19.7
17.1
15.3
13.9
12.9
12.1
11.4
10.8
10.3
9.8
9.5
9.1
8.8
8.5
8.3
8.0
7.8
7.6
7.4
7.3
7.1
7.0
6.8