Activate Acoustic Signals; (Beep); Select Attenuation (Fil); Switch Weight Unit (Unit) - Seca 703 User Manual

Hide thumbs Also See for 703:
Table of Contents

Advertisement

Activate acoustic
signals (BEEP)

Select attenuation (Fil)

Switch weight unit
52 •
You can select whether an acoustic signal is to be
emitted whenever a key is pressed or a stable weighing
value has been attained. The latter is relevant for the
Hold/Autohold function.
NOTE:
The function "Acoustic signal with stable weigh-
ing value" is activated at the factory. If required,
you can disable this function.
1. Select the item "BEEP" in the menu.
2. Confirm the selection.
3. Select a menu item:
– Press: Acoustic signal whenever a key is
– Hold: Acoustic signal with a stable weighing
4. Confirm your selection.
The current selection is displayed.
5. Select the setting you require:
– On
– Off
6. Confirm your selection.
You will leave the menu automatically.
7. If you also wish to activate the acoustic signal for
the second function, repeat this procedure.
You can use attenuation (Fil = filter) to reduce any inter-
ference during weighing (e.g. caused by patient mov-
ing).
1. Select the item "Fil" in the menu.
2. Confirm the selection.
The current selection is displayed.
3. Select an attenuation level.
– 0: low attenuation
– 1: moderate attenuation
– 2: high attenuation
4. Confirm the selection.
You will leave the menu automatically.
With uncalibrated scales you can select the unit you
(Unit)
wish to use for displaying the weight.
1. Select the item "Unit" in the menu.
pressed
value.