Error Messages

Error Messages
Error Messages
-101
Invalid character
An invalid character was found in the command string. You may have
inserted a character such as #, $, or % in the command header or within
a parameter.
For example, LIM:LOW O#.
-102
Syntax error
Invalid syntax was found in the command string.
For example, LIM:CLE:AUTO, 1 or LIM:CLE: AUTO 1.
-103
Invalid separator
An invalid separator was found in the command string. You may have
used a comma instead of a colon, semicolon, or blank space; or you may
have used a blank space instead of a comma.
For example, OUTP:ROSC,1.
-105
GET not allowed
A Group Execute Trigger (GET) is not allowed within a command string.
-108
Parameter not allowed
More parameters were received than expected for the command. You
may have entered an extra parameter, or added a parameter to a
command that does not accept a parameter.
For example, CAL 10.
-109
Missing parameter
Fewer parameters were received than expected for the command. You
omitted one or more parameters that are required for this command.
For example, AVER:COUN.
-112
Program mnemonic too long
A command header was received which contained more than the
maximum 12 characters allowed.
For example, SENSeAVERageCOUNt 8.