Power Meter Operation
Making Measurements using Sensor Calibration Tables
Note
Note
2-16
A calibration factor in the range of 1% to 150% can be entered.
The following rules apply to naming sensor calibration tables:
The name must consist of no more than 12 characters.
All characters must be upper or lower case alphabetic
characters, or numeric (0-9), or an underscore (_).
No other characters are allowed.
No spaces are allowed in the name.
2. To edit the parameter you currently have selected press
The power meter displays the data in a pop up window. Modify
this data (see below) as desired.
Use
or
is currently positioned.
Use
or
3. To confirm your choice press the appropriate softkey.
4. To add a new sensor calibration table entry press
prompted for the frequency and calibration factor. The entry is
sorted by frequency.
5. To remove a sensor calibration table entry use the
and
hardkeys to select the entry, press
delete the frequency the equivalent calibration factor is also
removed and vice versa.
You can only edit the table name you cannot delete it.
6. When you have completed editing the sensor calibration table
press
to return to the "Sensor Tables" screen.
Done
Ensure that the frequency points you use cover the frequency
range of the signals you want to measure. If you measure a signal
with a frequency outside the frequency range defined in the sensor
calibration table, the power meter uses the highest or lowest
frequency point in the sensor calibration table to calculate the
calibration factor.
to modify the character on which the cursor
to move to other characters.
HP EPM-441A User's Guide
.
Change
Insert
. You are
,
,
. If you