Error Messages
Error Messages
-330
Self-test Failed;Display Assy. Fault
Refer to "Display", on page 2-51 if you require a description of the
Display test.
-330
Self-test Failed;Confidence Check Fault
Refer to "Confidence Check", on page 2-48 if you require a description of
this test.
-350
Queue overflow
The error queue is full and another error has occurred which could not
be recorded.
-410
Query INTERRUPTED
A command was received which sends data to the output buffer, but the
output buffer contained data from a previous command (the previous
data is not overwritten). The output buffer is cleared when power has
been off, or after *RST (reset) command has been executed.
-420
Query UNTERMINATED
The power meter was addressed to talk (that is, to send data over the
interface) but a command has not been received which sends data to the
output buffer. For example you may have executed a CONFigure
command (which does not generate data) and then attempted to read
data from the remote interface.
-430
Query DEADLOCKED
A command was received which generates too much data to fit in the
output buffer and the input buffer is also full. Command execution
continues but data is lost.
-440
Query UNTERMINATED after indefinite response
The *IDN? command must be the last query command within a
command string.