Clear and Sanitize Procedures
Table 2: Nonvolatile Memory Devices (cont.)
Type and
minimum size
Function
NVRAM 128 kB
Instrument
configuration, option
keys, option usage
count
Flash ROM
GPIB configuration
128 kB
Flash ROM
Boot ROM
512 kB
ROM 22 kB
FPGA configuration
EEPROM 2 each
Temperature sensor
4 bytes
Microprocessor
Touch screen
– no details
controller
provided by
manufacturer
FLASH ROM
Display controller
128 kB
EPROM 8 kB
Front keypad
microcontroller
NVRAM 128 kB
OTP EPROM 2
Channel group
each 15 kB
microprocessor
EEPROM 2 each
Channel group
256 bytes
microprocessor
80E01, 80E02, 80E03, 80E04, 80E06, 80E07, 80E08, 80E09, 80E10, 80A01, 82A04, 80A05, 80A06 modules
EEPROM 2 kB
Module
configuration
memory
EEPROM 4 bytes
Temperature sensor
(2 each 80E07,
80E08, 80E09,
80E10)
80C01, 80C02, 80C03, 80C04, 80C05, 80C06, 80C07, 80C08, 80C09, 80C10, 80C11, 80C12 modules (including all options)
EEPROM 2 kB
Module
configuration
memory
EEPROM 4 bytes
Temperature sensor
2
User
Data input
1
modifiable
method
No
Programmed at
factory
No
Programmed at
factory
No
Programmed at
factory
No
Programmed at
factory
No
Not programmed
No
Programmed at
manufacturer
No
Programmed at
manufacturer
No
Programmed at
manufacturer
No
No
Programmed at
factory
No
Not used
Yes
Programmed
at factory;
also stores
calibration and
compensation
data for module
No
Not used
Yes
Programmed
at factory;
also stores
calibration and
compensation
data for module,
and user wave
lengths and
scales
No
Not used
Location
To clear
Processor
board
Processor
board
Processor
board
Processor
board
Processor
board
Display
adapter
board
Bridge board
Front panel
board
Acquisition
board
Acquisition
board
Acquisition
board
80Exx,
80Axx, 82Axx
main board
80Exx,
80Axx, 82Axx
sampler
boards
80Cxx main
board
80Cxx
sampler
boards
DSA8200 Declassification and Security
To sanitize
Not applicable – no user data
Not applicable – no user data
Not applicable – no user data
Not applicable – no user data
Not applicable – no user data
Not applicable – no user data
Not applicable – no user data
Not applicable – no user data
Not applicable – no user data
Not applicable – no user data
Not applicable – no user data
Not applicable – no user data
Not applicable – no user data
Use the sanitize procedure
documented below for user
wavelength(s) and scale(s)
Not applicable – no user data