Tektronix AFG2021 Technical Reference page 37

Arbitrary function generator specifications and performance verification
Hide thumbs Also See for AFG2021:
Table of Contents

Advertisement

Function
Amplitude
Sine
+ 4.0 dBm
+ 4.0 dBm
Sine
Sine
+ 4.0 dBm
Sine
+ 4.0 dBm
Sine
+ 4.0 dBm
Sine
+ 4.0 dBm
AFG2021 Arbitrary Function Generator Specifications and Performance Verification
e. Push the Amplitude/Level Menu bezel button.
f.
Push the -more- bezel button.
g. Push the Units bezel button and then push the dBm bezel button to change
the voltage units to dBm.
h. Push the -more- bezel button.
i.
Push the Amplitude bezel button (it will turn dark when activated) and
use the numeric keypad or general purpose knob to set the amplitude to
+4.0 dBm.
j.
Check that the Channel On/Off front panel button LED is lit.
If it is not lit, then the channel output is off. Push the Channel On/Off
button to turn it on.
3. Set the frequency of the Power Meter to 100 kHz.
4. Write the Power Meter reading of the 100 kHz sine wave as a reference
power value.
5. Verify that the power measurement at each frequency is within the error
specified in the following table.
Frequency
100.00 kHz
500.00 kHz
1.00 MHz
5.00 MHz
10.00 MHz
20.00 MHz
Measurement
(dB)
= Reference
Performance verification
Range
(dB)
- - - - - -
Reference ± 0.15
Reference ± 0.15
Reference ± 0.30
Reference ± 0.30
Reference ± 0.30
27