ZOOm:ZOOM<x>:<wfm>:VERTical:POSition
Description
This command sets or queries the vertical position of the specified wa-
veform for the specified zoom, where x is an integer from 1 to 4 repre-
senting the desired zoom window. The setting of the
ZOOm:ZOOM<x>:SCROLLLOCk command determines the wave-
forms that are affected.
Group
Zoom
Syntax 1
ZOOm:ZOOM<x>:<wfm>:VERTical:POSition <NR3>
Syntax 2
ZOOm:ZOOM<x>:<wfm>:VERTical:POSition?
Argument
H <NR3>
This is the vertical position, expressed in divisions.
Example 1
ZOOm:ZOOM1:CH1:VERTical:POSition 2
This command sets the Zoom1 vertical position to 2, which centers the
zoom trace at the second division of the acquired trace.
Example 2
ZOOm:ZOOM1:CH1:VERTical:POSition?
This query might return :ZOOm1:CH1:VERTical:POSition
0.0000, indicating that the Zoom1 trace for Channel 1 is centered at
division 0 of the acquired trace.
TDS5000B Online Programmer Guide