Measurement:meas<X>:Delay:direction - Tektronix TDS5000B Series Online Programmer Manual

Digital phosphor oscilloscopes
Hide thumbs Also See for TDS5000B Series:
Table of Contents

Advertisement

MEASUrement:MEAS<x>:DELay:DIREction
Description
This command sets or queries the starting point and direction that deter-
mines the delay "to" edge when taking a delay measurement. Use the
MEASUrement:MEAS<x>:SOURCE2 command to specify the wave-
form. This command is equivalent to selecting Time from the Measure
menu, choosing Delay from the drop- -down list and then clicking the
desired Search Direction setting. Measurements are specified by x,
which ranges from 1 through 8.
Group
Measurement
Related Commands
MEASUrement:MEAS<x>:SOURCE2
Syntax 1
MEASUrement:MEAS<x>:DELay:DIREction {BACK-
Wards|FORWards}
Syntax 2
MEASUrement:MEAS<x>:DELay:DIREction?
Arguments
H BACKWards
This means that the search starts at the end of the waveform and looks
for the last rising or falling edge in the waveform. Use the MEASUre-
ment:MEAS<x>:DELay:EDGE2 command to specify the slope of the
edge.
H FORWards
This means that the search starts at the beginning of the waveform and
looks for the first rising or falling edge in the waveform. Use the
MEASUrement:MEAS<x>:DELay:EDGE2 command to specify the
slope of the edge.
Example 1
MEASUrement:MEAS3:DELay:DIREction BACKWards
This command starts searching from the end of the waveform record.
Example 2
MEASUrement:MEAS3DELay:DIREction?
This query might return
:MEASUREMENT:MEAS3:DELAY:DIRECTION BACKWARDS
indicating that the current search direction is backwards.
TDS5000B Online Programmer Guide
-- 568 --