Trigger:a:holdoff:by - Tektronix TDS5000B Series Online Programmer Manual

Digital phosphor oscilloscopes
Hide thumbs Also See for TDS5000B Series:
Table of Contents

Advertisement

TRIGger:A:HOLDoff:BY

Description
This command sets or queries the type of holdoff for the A trigger.
Holdoff types are expressed as either user- -specified time (TIMe) or by
an internally calculated minimum time value (DEFAult/AUTO). This
command is equivalent to selecting Holdoff from the Trig menu and
then setting the Holdoff type.
Group
Trigger
Related Commands
TRIGger:A:HOLDoff:TIMe
Syntax 1
TRIGger:A:HOLDoff:BY {TIMe|DEFAult}
Syntax 2
TRIGger:A:HOLDoff:BY {TIMe|RANDOM|AUTO}
Syntax 3
TRIGger:A:HOLDoff:BY?
Arguments
H TIMe
This enables you to set the holdoff time via the TRIGger:A:HOLD-
off:TIMe command.
H DEFAult
This automatically calculates a holdoff time to use. This time is typi-
cally equivalent to the greater of 1/2 screen (5 divisions) of time or
250 ns. The maximum value is 12 s. For example, if the instrument is
set to 1 ms/division then the default holdoff will be 1 ms/division x 25
divisions = 25 ms.
Example 1
TRIGger:A:HOLDoff:BY?
This query might return
:TRIGGER:A:HOLDOFF:BY TIME
indicating that you will set the holdoff time.
Example 2
TRIGger:A:HOLDoff:BY TIMe
This command sets the holdoff to the "by time" setting. This enables
you to set the holdoff time.
TDS5000B Online Programmer Guide
-- 674 --

Advertisement

Table of Contents
loading

Table of Contents