DATa:ENCdg

Description
This command sets or queries the format of outgoing waveform data.
This command is equivalent to setting WFMOutpre:ENCdg,
WFMOutpre:BN_Fmt, and WFMOutpre:BYT_Or. Setting the
DATa:ENGdg value causes the corresponding WFMOutpre values to be
updated and visa versa.
Note: Values are constrained (for outbound data) to the format of the
data specified by DATa:SOUrce.
Group
Waveform Transfer
Related Commands
WFMOutpre:ENCdg, WFMOutpre:BN_Fmt, WFMOutpre:BYT_Or
Syntax 1
DATa:ENCdg {ASCIi|FAStest|RIBinary|RPBina-
ry|FPBinary|SRIbinary
|SRPbinary|SFPbinary}
Syntax 2
DATa:ENCdg?
Argument
H•ASCIi
This specifies the ASCII representation of signed INT, FLOAT. If
ASCII is the value, then :BN_Fmt and :BYT_Or are ignored.
H•FAStest
This specifies that the data be sent in the fastest possible manner con-
sistent with maintaining accuracy and is interpreted with respect to the
first waveform specified in the DATA:SOUrce list.
:ENCdg will always be BIN, :BYT_Or will always be LSB, but
:BN_Fmt and :BYT_Nr will depend on the first DATa:SOUrce wave-
form. :BN_Fmt will be RI unless the waveform is internally stored as
a floating point number, in which case the FP format will be used.
H•RIBinary
This specifies signed integer data point representation with the most
significant byte transferred first.
When :BYT_Nr is 1, the range is from - -128 through 127. When
:BYT_Nr is 2, the range is from - -32,768 through 32,767. When
:BYT_Nr is 8, then the waveform being queried has been set to Fast
Acquisition mode. Center screen is 0 (zero). The upper limit is one
division above the top of the screen and the lower limit is one division
below the bottom of the screen. This is the default argument.
TDS5000B Online Programmer Guide
-- 183 --