1
BB-4A Series Blackbody Calibrator
1.3
General Description
The Model BB4A is a portable, rugged, bench-top, blackbody calibration source
with a built-in precision PID digital controller. The calibrator is used to test and
calibrate infrared pyrometers. The heater cavity, with 21.6 mm (0.88") aperture,
has an emissivity of 0.99 and can be set to any temperature between 100 to 982°C
(212 to 1800°F).
1-2
IEC symbols
Description
Caution, risk of electric shock
Caution, refer to accompanying documents
115 VAC @50/60Hz (Domestic Models)
230 VAC @50/60Hz (European Models)
Figure 1. IEC symbols