Error Messages

MESSAGE
30
WHAT IT MEANS
Appears when the battery
cannot provide enough
power for a test.
Appears when a used test
strip is inserted
Appears when test strip is
removed while counting
down, or insufficient
blood volume.
Appears when ambient
temperature is below the
system operation range.
Appears when ambient
temperature is above the
system operation range.
Problem with the meter.
WHAT TO DO
Replace the battery immediately.
Repeat with a new test strip.
Re-test with a new test strip.
System operation range is 10°C to
40°C (50°F to 104°F). Repeat the
test after the meter and test strip
are in the above temperature range.
Repeat the test with a new test
strip.
If the meter still does not work,
please contact your customer
services for assistance.