Equinox Systems EQ-TD-4286 Owner's Manual page 18

Table of Contents

Advertisement

Out-of-range results
If you continue to have test results fall outside the range printed on the
test strip vial, the meter and strips may not be working properly. Do NOT
test your blood. Contact the local customer service or place of purchase
for help.
NOTE:
• The control solution range printed on the test strip vial is for control
solution use only. It is not a recommended range for your blood glucose level.
• See the Meter Maintenance section for important information about your
control solutions.
4. Read and compare the result
After counting down to 0, the control
solution test result will appear on the
display. Compare this result with the
range printed on the test strip vial and it
should fall within this range. If not, please
read the instructions again and repeat
109~165
the control solution test.
(136 mg/dL = 7.6 mmol/L; 109-165 mg/
dL = 6.1-9.2 mmol/L)
17

Advertisement

Table of Contents
loading

Table of Contents