WARNING:
When doing the control solution test, you have to mark it so that the test result
will NOT mix with the blood glucose TEST RESULTS stored in the memory. Failure
to do so will mix up the blood glucose test results with the control solution test
results in memory.
3. Apply control solution
Shake the control solution vial thoroughly before use. Squeeze out a
drop and wipe it off, then squeeze out another drop and place it on the
tip of the vial cap. Hold the meter to move the absorbent hole of the test
strip to touch the drop. Once the confirmation window fills completely,
the meter will begin counting down. To avoid contaminating the control
solution, do not directly apply control solution onto a strip.
16
2. Press M to mark this test as a control
solution test
With "
" displayed, the meter will
store your test result in memory under "
". If you press M again, the "
will disappear and this test is no longer a
control solution test.
"